System Error Codes | 289
Code Message What to do
1359 RAD Technology is not
supported
RAD single-tube and 8-to-1 cassettes are not
supported on this instrument. If the
hardware is properly installed and all cables
are currently plugged in, contact BioTek
TAC.
1401 The waste bottles should be
emptied before continuing
Empty the waste bottles and rerun the
protocol.
1404 Plate type restricted This instrument model or the requested
device does not support the selected plate
type. Edit the protocol to change the plate
type.
1405 Z-axis out of range The requested travel/dispense height cannot
be reached. The conflict may be caused by a
combination of variables, plate type, Plate
Clear or other height settings. Review the
protocol parameters and instrument settings
to identify a correction.
1407 Invalid step type The protocol may have been created for a
different instrument, it is not compatible
with this instrument.
1408 Invalid plate geometry Contact BioTek TAC.
1409 Invalid plate carrier type Contact BioTek TAC.
1412 Tip clearance error Contact BioTek TAC.
1413 AutoPrime in progress Contact BioTek TAC.
1414 AutoPrime aborted Contact BioTek TAC.
1415 AutoPrime value out of range Contact BioTek TAC.
1430 Strip washer syringe pump
FMEA
Contact BioTek TAC.
1431 Strip washer aspirate manifold
not installed
Contact BioTek TAC. Instrument basecode is
not correctly configured.
1432 Strip washer module not
connected
Plug in the strip washer: connect the washer
module to the MultiFlo FX using the supplied
serial cable.
1433 Bad step pointer used to
calculate Z-axis positioning
Contact BioTek TAC.
BioTek Instruments, Inc.