290 | Appendix A: Error Codes
Code Message What to do
1440 RAD module not connected. Check the cables that connect the external
Peri-pump and the RADmodule to the back
of the instrument. Make sure they are
correctly plugged in.
1441 RAD 8-to-1 bulk-dispensing
cassette Chute Height invalid.
Change the Peri-pump advanced instrument
setting for Chute Height.
1503 Dispense volume error Contact BioTek TAC.
1506 The requested carrier Y-axis
position is out of range
Contact BioTek TAC.
1508 Pre-dispense volume error The specified Pre-dispense volume is invalid
for this plate type. Edit the protocol.
1512 Invalid plate type The specified plate type is not supported
with the current hardware.
1513 Plate type - manifold conflict Change the Plate Type to one supported by
the manifold. Or, check the Instrument
Settings and make sure they match the
physical hardware: Get settings from
instrument.
1600-
160C
The onboard storage space
allocated for this function has
been used up
Use the LHC Manage Memory control to
reallocate space.
160D Not a valid step Contact BioTek TAC.
2400 Parameter limit exceeded Contact BioTek TAC.
4000 Program locked so operation
denied
Contact BioTek TAC.
4010 Program cannot be erased so
delete denied
Contact BioTek TAC.
4020 Bad checksum when reading
program from EEPROM
Contact BioTek TAC.
4030 Program not found Contact BioTek TAC.
4040 Cant save program because no
space available
Contact BioTek TAC.
4050 Program run canceled by user Restart instrument if this message is
unexpected.