System Error Codes | 291
Code Message What to do
8100 Communications NAK Contact BioTek TAC.
In this 8100 series of communications error codes, the third number, e.g.
81n1, n identifies the source: 0 between PC and instrument; 1 between UI and
MC; and 2 between BioStack and instrument. For example, 8101 is timeout
error in communication between the instrument and the PC, and 8111 is a
timeout error in communication between the internal controllers, UI and MC.
8101
8111
Timeout while waiting for serial
message data
Contact BioTek TAC.
8102 Instrument busy and unable to
process message
Contact BioTek TAC.
8103 Receive buffer overflow error Contact BioTek TAC.
8104 Checksum error Contact BioTek TAC.
8105 Invalid structure type in
byMsgStructure header field
Contact BioTek TAC.
8106 Invalid destination in
byMsgDestination header field
Contact BioTek TAC.
8107 Request object received not
supported by instrument
Contact BioTek TAC.
8108 Message Body size exceeds
max limit
Contact BioTek TAC.
8109 Max number of requests
currently running and cannot
run the latest request
Contact BioTek TAC.
810A No request running when
response request issued
Contact BioTek TAC.
810C Response for outstanding
request not ready yet
Contact BioTek TAC.
810D To communicate with the LHC,
the instrument must be at its
main menu
The LHC can only talk to the instrument
when its main menu is displayed. When the
instrument is busy, for example when
AutoPrime is running, press the Stop button
on the keypad, if desired, to end the current
process and return to the main menu.
810E One or more request
parameters are not valid
Contact BioTek TAC.