ICS-3000 Ion Chromatography System
130 Doc. 065031-04 1/08
8. Enter the Minimum Pressure and Maximum Pressure values under
Pressure Display on the Control panel. Setting high and low pressure limits
ensures that the DP/SP will automatically stop if a system malfunction occurs.
Notes
When the system includes an EG, the high pressure limit for the DP/SP is
21 MPa (3000 psi) and the low pressure limit is 1.4 MPa (200 psi).
The DP/SP is equipped with a pressure alarm limit that prevents operation
above 34 MPa (5000 psi).