EasyManua.ls Logo

Dionex ICS-3000 - Page 200

Dionex ICS-3000
392 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ICS-3000 Ion Chromatography System
184 Doc. 065031-04 1/08
Injection valve is not actuating or the sample loop is plugged
1. Test the valve by manually switching the position from the
Detector/Chromatography tab on the panel tabset.
2. Check the sample loop for blockage. Clean or replace the loop if needed.
Detector not properly installed
CD: An electronics connector on the back of the detector plugs into a
receptacle on the detector compartment. Push on the detector to make certain
the connector is securely connected (see Figure 8-1
).
ED: Verify that the signal cables from the cell are connected to the detector
block (see Figure 8-2
). In addition, an electronics connector on the back of the
detector plugs into a receptacle on the detector compartment. Push on the
detector to make certain the connector is securely connected.
Figure 8-1. Conductivity Detector Connection
Figure 8-2. Electrochemical Detector Connection
Push here to ensure that the detectors
electrical connection is secure.
Push here to ensure that the detector’s
electrical connection is secure.

Table of Contents

Related product manuals