ICS-3000 Ion Chromatography System
194 Doc. 065031-04 1/08
This error occurs if you attempt to use relay 4 as a standard relay in a
Chromeleon or Chromeleon Xpress program, although the relay has been set
to synchronize the gradient with the autosampler sample injection.
To troubleshoot:
Define the relay as a standard relay or specify a different relay for use in the
program.
If the DP/SP pressure exceeds 5 MPa (725 psi) during priming, this error
occurs and the priming process is aborted.
To troubleshoot:
1. Verify that the priming valve is open and then reprime the pump. (To open
the valve, turn the knob one-half turn counterclockwise.)
2. Check fittings for overtightening. Check tubing for crimping and replace
it as necessary. Be careful not to overtighten fittings.
This error may be caused by a blockage in the system or by a problem with
the column. (“Right-hand” refers to the bottom pump; “left-hand” refers to the
top pump in a dual-pump module.)
Relay 4 is configured for inject synchronization. Please
change pump configuration.
The maximum purge pressure was exceeded.
The pressure in the left-hand working cylinder exceeded the
safety limit.
-or-
The pressure in the right-hand working cylinder exceeded the
safety limit.
-or-
The system pressure exceeded the safety limit.