ICS-3000 Ion Chromatography System
206 Doc. 065031-04 1/08
To troubleshoot:
1. Connect the EluGen cartridge cable connection to the electrical bulkhead
(see Figure 2-8
).
2. Refer to Section 8.23
to determine why there is no flow.
3. If the error message appears again, contact Dionex for assistance. The
cartridge control electronics may have malfunctioned.
NOTE The ICS-3000 Ion Chromatography System electronics
components cannot be serviced by the user.
This error occurs when, in order to maintain the selected current, the power
supply is required to supply a higher voltage than the EluGen cartridge can
support.
To troubleshoot:
Replace the EluGen cartridge (see Section 9.12
).
This error occurs when the current applied to the EluGen cartridge exceeds
the maximum current allowed. (The EluGen current is automatically turned
off to prevent damage to the cartridge.) This error may also occur if the liquid
flow to the cartridge is interrupted.
To troubleshoot:
1. Connect the EluGen cartridge cable connection to the electrical bulkhead
(see Figure 2-8
).
2. Refer to Section 8.23
to determine why there is no flow.
EG1 over power.
-or-
EG2 over power.
EG1 over voltage.
-or-
EG2 over voltage.