8 • EG Troubleshooting
Doc. 065031-04 1/08 205
This error occurs if the flow rate is set to a value the EG does not support. The
DP/SP flow rate range is 0.001 to 10.0 mL/min; however, when an EG is
installed, the allowed range is 0.01 to 3.00 mL/min. The recommended
operating range is 0.4 to 2.00 mL/min.
To troubleshoot:
Set the flow rate to a value within the allowed range.
This error occurs if the selected concentration is too high for the current flow
rate.
To troubleshoot:
Set the flow rate to a value within the allowed range. The allowable eluent
concentration for a particular application depends on several factors: the flow
rate, suppressor type, EluGen cartridge type, and cartridge configuration. For
details, refer to Section 4.7
.
This error occurs when the current applied to the EluGen cartridge exceeds
the maximum current allowed. (The EluGen current is automatically turned
off to prevent damage to the cartridge.) This error may also occur if the liquid
flow to the cartridge is interrupted.
EG1 invalid flow.
-or-
EG2 invalid flow.
EG1 invalid flow rate-concentration.
-or-
EG2 invalid flow rate-concentration.
EG1 over current.
-or-
EG2 over current.