ICS-3000 Ion Chromatography System
228 Doc. 065031-04 1/08
When a TC is shared by two timebases, the same values must be selected for
certain parameters. If the Chromeleon Ready Check detects a discrepancy,
this error occurs and batch processing cannot be started.
To troubleshoot:
Find and eliminate any discrepancies in the parameter settings. For example,
all samples must turn on TempCtrl or all samples must turn off Temp Ctrl.
Selecting the Disconnect or Standby command during data acquisition will
stop data transmission and cause data acquisition to stall. If you attempt to
issue one of these commands during data acquisition, this error occurs.
To troubleshoot:
Wait until data acquisition is completed before attempting to issue either of
these commands.
All samples scheduled for run on timebases x need to specify
the same value for property (nominal) Temperature.
-or-
All samples scheduled for run on timebases x need to specify
the same value for property TempCtrl.
-or-
All samples scheduled for run on timebases x need to specify
the same value for property y.
Can't execute “Disconnect” command during data
acquisition.
-or-
Can't execute “Standby” command during data acquisition.