ICS-3000 Ion Chromatography System
232 Doc. 065031-04 1/08
This error occurs if you attempt to issue a command that is not available when
the TC is in demo (or virtual) mode.
To troubleshoot:
If a TC is present and connected, disable the demo mode:
1. Start the Server Configuration program.
2. In the timebase, right-click the TC and select Properties on the context
menu.
3. On the General tab page in the Properties dialog box, clear the Demo
Mode check box and click OK.
This error occurs if there is a communication failure between the TC and the
PC on which Chromeleon or Chromeleon Xpress is installed.
To troubleshoot:
1. Check the USB connections.
2. Check the connection from the TC to the power supply.
3. If you cannot identify the cause of the error, contact Dionex for
assistance.
This error occurs if a TC firmware (Moduleware) command is not executed
within the time allotted for its completion.
To troubleshoot:
1. Check the USB connections.
2. Check the connection from the TC to the power supply.
3. If you cannot identify the cause of the error, contact Dionex for
assistance.
Command is not available in demo mode.
Communication error: x.
Communication time-out.