ICS-3000 Ion Chromatography System
238 Doc. 065031-04 1/08
This error occurs if the gas sensor detects an increased concentration of gas
inside the TC.
To troubleshoot:
1. Find and eliminate the leak.
2. On the TC Control panel, click the Clear Alarm button.
3. Ventilate the interior of the TC thoroughly before closing the door and
resuming operation.
This error occurs if the humidity sensor detects an increased concentration of
humidity inside the TC.
To troubleshoot:
1. Find and eliminate the leak.
2. On the TC Control panel, click the Clear Alarm button.
3. Ventilate the interior of the TC thoroughly before closing the door and
resuming operation.
This error occurs if you enter a parameter that is not valid for the TC.
To troubleshoot:
Select a valid parameter.
Gas leak detected.
Humidity leak detected.
Illegal parameter.