240 Doc. 065031-04 1/08
The Qualification.WarningPeriod property determines the number of days
before the due date that the warning is first displayed. The
Qualification.GracePeriod property determines for how many days after the
due date you can continue using the TC without performing the qualification.
To troubleshoot:
Contact Dionex to schedule an Operational Qualification and Performance
Qualification.
Chromeleon and Chromeleon Xpress monitor the date on which service was
last performed and display an appropriate message (shown above) at the
following times: when the due date for the next service is approaching and
when the due date is overdue. The Service.WarningPeriod property
determines the number of days before the due date that the warning is first
displayed.
To troubleshoot:
Contact Dionex for information about performing annual preventive
maintenance. The DC/TC Preventive Maintenance Kit (P/N 061796) contains
all required replacement parts.
Next service of this module is due in x day(s) (due date is
[date]).