GE HEALTHCARE
DIRECTION 5394152, Revision 5
LOGIQ™ P6/P6 PRO SERVICE MANUAL
Chapter 4 - Page 4-15
4-3-10-1 Using the DVD-R (cont’d)
15.)It will do Verifying.
16.)It will eject the CD/DVD and Go to Utility and reinsert the CD/DVD.
17.)Go to Utility->Connectivity->Removable Media.
18.)Press Verify
19.)The properties will be like below.
a.) Capacity is the size of transferred data.
b.) Free space is 0.
c.) Formatted / Database Present / Finalized : Yes
(DICOMDIR Present : depends on setting)
Figure 4-11 Verifying
Figure 4-12 Removable Media
NOTICE
Avoid mechanical ejection whenever possible. Mechanical ejection leaves the actuator
unlocked. If forced to use this method, reboot the system, then insert and eject a known good
disk using one of the other methods.