GE HEALTHCARE
DIRECTION 5394152, Revision 5
LOGIQ™ P6/P6 PRO SERVICE MANUAL
Chapter 8 - Page 8-111
Section 8-4Software Installation
8-4-1 Installing Base Image Software
NOTE: While we do not believe it is necessary, It would not hurt to disconnect the system from the
network and remove all transducers.
NOTE: All digital peripherals, reporting systems connected via the USB ports should be either power
off or disconnected prior to the software installation
1.) Power ON the scanner.
2.) Tap gently the DEL key until the following screen appears.Enter “arirang” as a password if the
system request a password, then press the Enter key.
3.) Move to Advanced and select IDE Configuration.
Figure 8-163 Enter Current Password
Figure 8-164 IDE Configuration