LOGIQ™ P6/P6 PRO SERVICE MANUAL
Page 8-96 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-44-3 Removal procedure (cont’d)
4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-44-4 Mounting Procedure
Install the new parts in the reverse order of removal.
Table 8-45 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test Debrief Script
Section 4-3-1
Power On/Boot Up
“Service Manual, Direction
5394152, Rev 1+, Section 8-2-44. Equipment
passed all required tests and is ready for use. “
Section 4-3-2
Power Off / Shutdown
Section 4-9-13
ACPC II Assy set Function Validation
Procedure
Section 10-5-2-2
Peripheral/Option Checks
Section 10-5-5
Probe Maintenance
Section 10-6-2
GEMS Leakage Current Limits
Section 10-6-3
Outlet Test - Wiring Arrangement - USA &
Canada
Section 10-6-4
Grounding Continuity
Section 10-6-5
Chassis Leakage Current Test
Section 10-6-6
Isolated Patient Lead (Source) Leakage–Lead
to Ground
Section 10-6-7
Isolated Patient Lead (Source) Leakage–Lead
to Lead
Section 10-6-9
Probe Leakage Current Test