GE HEALTHCARE
DIRECTION 5394152, Revision 5
LOGIQ™ P6/P6 PRO SERVICE MANUAL
Chapter 5 - Page 5-29
5-7-6 Diagnostics
NOTE: When Service Login for FRU Diagnostic, scan mode must be B-mode.
Detailed Diagnostic information is found in Diagnostics/Troubleshooting.
NOTE: After finishing Diagnostic test, the system should be rebooted.
5-7-6-1 Diagnostics Execution
Diagnostic tests are executable by both local and remote users. The Service Platform provides top-level
diagnostic selection based on the user’s level and login access permissions. Remote access will require
disruptive diagnostic permissions to run Acquisition diagnostics.
5-7-6-2 Diagnostic Reports
Diagnostic tests return a report to the Service Platform. The platform retains the report and allows for
future viewing of the diagnostic logs.
5-7-6-3 Proactive Diagnostics
A system of self-monitoring is largely supported with the integration of Insite2. The scheduler, executive,
user interfaces, and some of the proactive diagnostic functions are provided by Insite2. Other tasks
should be provided by the Product team.
WARNINGWARNING
Diagnostics should be performed with probes which have 128 or more elements
Figure 5-28 GE Service Diagnostic Page