GE HEALTHCARE
DIRECTION 5394152, Revision 5
LOGIQ™ P6/P6 PRO SERVICE MANUAL
Chapter 8 - Page 8-21
8-2-4-4 Assembly procedure
1.) Insert two of the flexible arm stoppers (A+B) to the Power Cable and DVI cable in the Order
illustrated on the figure below. Make sure that chamfered shape of the flexible Arm stopper should
face inside.
2.) Insert oilless bush into the flexible arm neck pipe and insert two cables into the flexible arm neck
pipe.
3.) Insert the flexible arm into the flexible arm neck pipe.
Figure 8-36 Inserting cables to the flexible arm stopper
Figure 8-37 Inserting the oilless bush
Figure 8-38 Inserting the flexible arm
Flexible Arm Stopper