GE HEALTHCARE
DIRECTION 5394152, Revision 5
LOGIQ™ P6/P6 PRO SERVICE MANUAL
Chapter 8 - Page 8-61
8-2-23 Dummy Cover L / R
8-2-23-1 Tools
Common Phillips screwdrivers
8-2-23-2 Preparations
Shut down the system and switch off the main breaker.
8-2-23-3 Removal procedure
1.) Remove 2 rubber caps (1-2).
2.) Unscrew 2 screws and remove the dummy cover L.
3.) Same process is applied for dummy cover R.
Figure 8-100 rubber cap of dummy cover
Figure 8-101 screws of dummy cover

Table of Contents