LOGIQ™ P6/P6 PRO SERVICE MANUAL
Chapter 10 - Page 10-19
10-6-5-4 Data Sheet for Chassis Source Leakage Current
The test passes when all readings measure less than the value shown in Table 10-14. Record all data
on the PM Inspection Certificate.
Table 10-18 Typical Data Sheet for Chassis Source Leakage Current
Unit Power
Tester Polarity
Switch
Tester Neutral
or Ground
Switch
Test 1
Probe
Connector
Test 2
caster
Test 3
CRT
Optional
Test 4
Optional
Test 5
Enter Name of tested peripheral here:
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
OFF NORM OPEN
OFF NORM CLOSED
OFF REV OPEN
OFF REV CLOSED