8-2-26-3 Removal procedure (cont’d)
6.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-26-4 Mounting Procedure
Install the new parts in the reverse order of removal.
Table 8-27 Functional Tests
Service Manual
Section Functional Test / Diagnostic Test Debrief Script
Section 4-3-1
Power On/Boot Up
“Service Manual, Direction
5394152, Rev 1+, Section 8-2-26. Equipment
passed all required tests and is ready for use. “
Section 4-3-2
Power Off / Shutdown
Section 4-9-2
DTRX64II Assy Function Validation Procedure
Section 4-3-4
System B/M-Mode Checks
Section 4-3-5
System CFM and PWD Checks
Section 10-5-2
Functional Checks (See Also Chapter 4)