GE HEALTHCARE
DIRECTION 5394152, Revision 5
LOGIQ™ P6/P6 PRO SERVICE MANUAL
Chapter 4 - Page 4-43
4-9-4 CWD64 Assy Function Validation Procedure (cont’d)
4.) Press “CW” button to start CWD Mode.
5.) Scan in CWD mode (with Virtual Convex enabled).
6.) Press “B” button to scan in “B” mode.
7.) Press Virtual Convex button to disable Virtual Convex if not already disabled.
8.) Press “CW” button to start CWD Mode.
9.) Scan in CWD mode (with Virtual Convex disabled)
10.)Compare the CWD (with Virtual Convex enabled) sensitivity and CWD (with Virtual Convex
disabled).
11.)Pass if two modes have the same sensitivity.
12.)Do the service login. Refer to the 5-7-3-2 "Service Login" on page 5-25.
13.)Select “Diagnostics” => “LOGIQ P6 Diagnostics” => “FRU Test” => “CWD64” => “CWD64 Overall”
NOTE: For LP6 PRO, select 'LOGIQ P6 PRO Diagnostics' => 'FRU test' => 'ACWD' => “ACWD
Overall”.
Figure 4-41 CW mode
Figure 4-42 Virtual Convex Disabled.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P6 Series and is the answer not in the manual?

GE LOGIQ P6 Series Specifications

General IconGeneral
Power Supply100-240 VAC, 50/60 Hz
Imaging ModesB-Mode, M-Mode, Color Doppler, Power Doppler
ApplicationsVascular, Musculoskeletal
Operating Temperature10°C to 40°C
Display15-inch LCD
Relative Humidity20% to 80% (non-condensing)

Related product manuals