GE HEALTHCARE
DIRECTION 5394152, Revision 5
LOGIQ™ P6/P6 PRO SERVICE MANUAL
Chapter 8 - Page 8-109
8-3-5-2 Image Management Guide (cont’d)
Importing Data
To import an exam(s) to another Ultrasound system:
1.) At the other Ultrasound system, insert the CD-ROM/DVD.
2.) Press Patient. Press Exam Data Transfer (located at the upper, left-hand corner of the Patient
menu).
3.) Select Import task Button.
Figure 8-160 Exam Data Transfer
Figure 8-161 Task button