Page 8-150 Section 8-4 - Software Installation
8-6-2-2 Software Reloading (cont’d)
4.) When installation is finished, the system shuts down automatically.
5.) Press Power Button to start the system.
6.) When system fully boots up after the software reloading, please shutdown system again.
NOTE: For the following procedure, disruptive mode should be approved in order to be in done
remotely.
7.) Start the system.
8.) Press “Patient” button, and verify under Dataflow that “Local Archive” is available. If it's not
available, shut down the unit one more time.
9.) To put the Local Time Zone back, Go to Utility->System->General
10.)Click “Date/Time”
Figure 8-226 Date/Time