GE HEALTHCARE
DIRECTION 5394152, Revision 5
LOGIQ™ P6/P6 PRO SERVICE MANUAL
Page 8-152 Section 8-4 - Software Installation
8-6-2-2 Software Reloading (cont’d)
15.)Change the Date and Time correctly if needed.
16.)Press “Apply”
17.)Press “OK”
18.)Turn off the system.
19.)Start the system.
20.)Check if the correct date and time are displayed.
21.)The software reloading is completed.
Figure 8-229 Date and Time check