Illustration 14: Left Rear Plug-in Panel
2. Plug the power connection for the cardiac trigger monitor into the electrical connection port
(4).
3. Connect the coaxial cable to the connection port above the heart (5).
NOTE: Do not loop the cable. This could introduce noise into the ECG trace.
4. Connect the Ethernet cable from the cardiac trigger monitor to the port labeled
Monitor
(6).
5. On the Gantry Display
Menu Area
, select Settings → ECG Settings to select the External
ECG as the primary cardiac device. The External ECG can also be selected from the
Console user interface.
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
Chapter 8 Equipment 177