Illustration 14: Coverage Speed collection — Axial
Table 13: Coverage Speed — Axial
Element Description
Table Positions
One
(Single)
One or More
(Multiple)
Use when the scan range is greater than 160 mm if using Smart Coverage or when
the fixed coverage scan range is greater than the detector coverage.
Detector Coverage
Smart Coverage (system adjusts to best coverage based on the start and end loca‐
tion prescribed)
5.0 mm
40 mm
80 mm
120 mm
140 mm
160 mm
Also see the
Equipment
chapter;
Detector Configuration
section for more information.
Number of Passes
The number of rotations for the data acquisition. Only available if the interval is 0.
Minimum Time Between Passes
The time between rotations for the data acquisition when the interval is 0.
Scan Interval
0, detector coverage or greater than detector coverage
Rotation Time
0.28 s, 0.35 s, 0.5 s, 1.0 s
Total Exposure Time
The coverage time determines the total X-ray On duration of the scan.