1. Section 2.4, Open EA3.
2. From the
Application
tab, select the desired parameters.
Illustration 14: Application Tab
Field Description
Enable Authorization
Enable or disable authorization.
If authorization is enabled, anyone logging in through EA3 (both local and enterprise users)
must have a role. Anyone without a role is denied access.
The user role does not matter for logging in through EA3, however, other EA3 client applica‐
tions may restrict which roles can login.
Emergency Logon Allowed
Enable or disable emergency access.
If EA3 is used in GUI mode, this determines whether or not the [Emergency Login] button dis‐
plays.
If disabled, emergency user access is prevented.
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
606 2 Access Controls