Quantitative Analysis
9-18 Vivid S70 / S60User Manual
BC092760-1EN
01
Trace smoothing
The system can smooth the traces displayed by applying a filter
over a defined time window. The type of filter available is
depending on the analysis signal displayed.
1. If necessary, press Trackball until the QA trackball
assignment is selected.
2. With the cursor in the Analysis window, press Update
menu.
The System menu is displayed.
3. Select Smoothing.
The Smoothing menu is displayed.
4. Select a smoothing filter.
The trace display is updated.
Cine compound
Cine compound calculates and displays cineloops generated
from a temporal averaging of multiple consecutive heart cycles.
The number of averaged cycles is displayed on the top left
corner.
To apply cine compound:
1. Adjust Cine compound to set the number of heart cycles to
average.
The traces are updated showing averaged data. The
number of heart cycles averaged is displayed on the top left
corner.
2. Press CC Zoom to display the last recorded heart cycle.
3. Press CC Zoom again to unzoom.