Care and Maintenance
Vivid S70 / S60 – User Manual 13-21
BC092760-1EN 01
Probe Manual Cleaning Instructions (Required for All Probes)
1. Put on a clean pair of gloves and fill a sink or basin with
warm potable water (30 - 40°C) to a level allowing
immersion of the probe up to the immersion line shown in
Figure 13-5.
2. Prepare the cleaning solution in accordance with the
disinfectant manufacturer’s instructions.
3. Immerse the probe in the prepared cleaning solution up to
the immersion line and ensure no air bubbles are trapped on
the surface. Do not submerge probe beyond the immersion
line shown in Figure 13-5.
Figure 13-5. Probe immersion levels
NOTE: For probe immersion levels for 6Tc-RS, 9T-RS, 6VT-D and
10T-D please refer to the user documentation enclosed with
the respective probes.
4. Brushing with a clean, soft, nylon bristle brush from the base
of the cable strain relief to the distal tip is critical to ensure
cleaning and disinfection efficacy.
a. M5Sc-D, 3Sc-RS6S-D, 12S-D
b. C1-5-D, C1-6-D, C2-9-D, C3-10-D
c. 9L-D, 11L-D
d. 2D (P2D)
e. 6D (P6D)
f. iC5-9-D
g. L8-18i-D
1. Fluid level