8. Visually inspect the device in a well-lit area to ensure all
surfaces are free from residual cleaning solution. Repeat
Step 7 if visible cleaning solution is observed.
9. Thoroughly dry the probe using a clean lint-free soft and dry
cloth or wipe.
Cable and Connector Manual Cleaning
1. The cable and connector surfaces can be cleaned with the
cleaners or wipes listed in the Probe Care Card.
NOTE: Use of wipes listed in the Probe Care Card may result in
discoloration of the cable.
2. Wipe the cable with a lint-free cloth dampened with potable
water to remove chemical residue.
Figure 13-8. Cleaning the Probe Cable
Use caution when cleaning the connector. This cable connector
should only be cleaned with a slightly dampened cloth or wipe.
Exposure to excessive moisture will result in damage to the
probe and possibly the ultrasound console. DO NOT wet the
connector/console interface surface or labels.