Probes
13-24 Vivid S70 / S60 – User Manual
BC092760-1EN
01
Probe Intermediate-Level Disinfection - Spray
NOTE: For intermediate-level disinfection of surface contacting probes,
choose either the spray or wipe method.
NOTE: High-level disinfection is required for devices that contact intact
mucous membranes or non-intact skin.
1. Put on a new pair of gloves and spray enough disinfectant
solution to saturate a new disposable lint-free wipe or cloth.
2. Holding the probe near the strain relief, apply the dampened
cloth to the patient contacting lens. Wipe the probe from the
lens to the strain relief, slightly rotating the probe after each
wiping pass.
3. After the probe has been completely wiped, dampen a
second wipe with disinfectant and starting at the probe lens
begin wiping the probe in a rotating motion moving down
towards the strain relief. Spray disinfectant directly on the
recessed areas and ridges to saturate.
Figure 13-9. Disinfecting the Probe Moving from Lens to Strain
Relief
4. Once the probe has been completely wiped, dampen a third
wipe with disinfectant and continue wiping the probe as
needed to ensure the surface remains wet for the required
exposure time. Use as many wipes as needed and respray
disinfectant on recessed areas and ridges, to ensure all
surfaces remain wet for the minimum required contact time
listed in the disinfectant manufacturer’s instructions for use.
5. Dry all surfaces of the probe using a sterile lint-free wipe or
cloth.
6. Saturate a sterile lint-free wipe with de-ionized or sterile
water (remove excess water, wipe should be damp but not