Care and Maintenance
Vivid S70 / S60 – User Manual 13-29
BC092760-1EN 01
Figure 13-12. Probe suspended in disinfectant basin
NOTE: When performing high-level disinfection of GE ultrasound
probes with the Trophon® EPR, it is not necessary to
disconnect the probe from the ultrasound system. The
probe must be inactive (not selected) during the disinfection
cycle.
3. Thoroughly rinse the probe by immersing it in a large
volume of critical (purified) water for a minimum of 1 (one)
minute. Remove the probe and discard the rinse water.
Do not reuse the water. Always use fresh volumes of water
for each rinse.Repeat Step 3 two additional times, for a total
of 3 (three) rinses.
4. Thoroughly dry all surfaces of the probe using a sterile,
lint-free wipe or cloth, changing wipes’ cloths when
necessary to ensure the probe is completely dry. Visually
inspect the probe to ensure all surfaces are clean and dry.
Repeat drying steps if any moisture is visible.
5. After each use, inspect the lens, cable, and housing of the
probe. Look for any damage that would allow liquid to enter
the probe. If the probe is damaged, do not place it into any
liquid (e.g. for disinfection) and do not use it until it has been
inspected and repaired/replaced by a GE Service
Representative.
Ensure that the probe is suspended. The probe face should not
be resting against the tank/basin surface and should be in full
contact with the liquid. Carefully place the probe in the basin,
taking care not to damage the transducer lens.
Failure to properly rinse probes with water following
disinfection may cause skin irritation.