Probes
13-30 Vivid S70 / S60 – User Manual
BC092760-1EN
01
The instructions provided above have been validated to properly
prepare GE Ultrasound probes for re-use. It remains the
responsibility of the processor to ensure that the processing, is
performed as specified in this document. This may require
verification and routine monitoring of the process.
Chemicals Used for Efficacy Validation
The table below lists the products and intended use (clean, ILD,
HLD) that were validated.
Table 13-4: Chemicals used for Efficacy Validation on non-TEE probes
Product
Type Trade Name Manufacturer
Minimum
Contact
Time
Active
Ingredient
Validated for
probe
Enzymatic
Detergent
(Soak)
Enzol®
(Cidezyme®)
Advanced
Sterilization
Products®
(J&J)
1-Minute
Soak
Proteolytic
Enzymes
3Sc-RS,
M5Sc-D, 6S-D,
12S-D, 9L-D,
11L-D, C1-5-D,
C1-6-D, C2-9-D,
C3-10-D,
iC5-9-D,
L8-18i-D
MetriZyme™ Metrex™
Prolystica® 2X
Concentrate
Presoak &
Cleaner
Steris
Intermediate-l
evel
Disinfectant
(wipe)
Oxivir® Tb Diversey 10-Minute
Exposure
Hydrogen
Peroxide
3Sc-RS,
M5Sc-D, 9L-D,
11L-D, C1-5-D,
C1-6-D, C2-9-D,
C3-10-D,
iC5-9-D,
L8-18i-D
High-level
Disinfectant
(Soak)
Cidex® OPA Advanced
Sterilization
Products (J&J)
10-Minute
Soak
Ortho-
phthalaldehyde
3Sc-RS,
M5Sc-D, 6S-D,
12S-D, 9L-D,
11L-D, C1-5-D,
C1-6-D, C2-9-D,
C3-10-D,
iC5-9-D,
L8-18i-D
McKessen
OPA/28