System Care and Maintenance
Vivid S70 / S60 – User Manual 15-5
BC092760-1EN 01
Compatible chemicals for cleaning
NOTE: Effective cleaning for parts with narrow gaps and holes (e.g.
keyboard, trackball...) is difficult.
NOTE: The system cabinet includes the probe holders.
Cleaning agents
Main display (glass)
Touch display (glass)
Footswitch
Main display frame (panel)
System cabinet
Operator controls
Mild, Non-Abrasive Soap and Water XXXXXX
Ammonia (2:1 ratio of max 10% Home Ammonia) X X X
Disinfectants
Bleach (10:1 ratio of 5% Home Bleach) X X X
Hydrogen Peroxide / Hydrogen Peroxide Wipes X X X
T-Spray II X
Sani-Cloth HB X
Diluted Cidex X
Cidex OPA X
Metricide 14 X
Wavicide-01 X
Isopropyl Alcohol (70%) XXXXX
Do not use T-Spray I (Original T-Spray) or SaniCloth Plus (red
cap).
System surfaces and transducers are resistant to ultrasound
gel, alcohol, and disinfectants, but if used, wipe them off to
prevent permanent damage.