Maintenance
15-6 Vivid S70 / S60 – User Manual
BC092760-1EN
01
Cleaning procedures
NOTE: The following procedures describe cleaning of the ultrasound
system components to a general housekeeping level only.
Prior to cleaning any part of the system turn off the power. If
possible, disconnect the power cord.
Main display and Touch screen
To clean the main display frame
1. Apply cleaning agent to a soft, non-abrasive folded cloth.
The cloth should be damp, not dripping wet.
2. Wipe down the top, front, back, and both sides of the display
frame. Do not spray any liquid directly onto the screen.
3. Wipe off excess cleaning agent and allow to dry.
To clean main display and touch panel (glass):
1. Apply cleaning agent to a soft, non-abrasive folded cloth.
The cloth should be damp, not dripping wet.
2. Gently wipe the surface of the main display and touch panel.
3. Wipe off excess cleaning agent and allow to dry.
Control panel and keyboard
To clean the Control panel and keyboard:
1. Apply cleaning agent to a soft, non-abrasive folded cloth.
The cloth should be damp, not dripping wet.
2. Gently wipe the surface of the control panel.
3. Gently wipe the surface of the keycaps.
4. Use a cotton swab to clean around keys or controls. Use a
toothpick to remove solids from between keys and controls.
When cleaning the operator control panel, make sure not to
spill or spray any liquid on the controls, into the system
cabinet, or in the probe connection receptacle.
To clean the trackball:
Clean the trackball only when required.
1. Rotate the top locking ring counter-clockwise around the
trackball.
2. Lift off the top locking ring, the rubber dust filtering ring and
the trackball from the Control panel.
3. Wipe the trackball with a soft dry cloth.
4. Wipe the trackball housing, rollers, etc. with a soft dry cloth
or a cotton swab.