User Manual corpuls
3
Configuration
ENG - Version 2.1 – P/N 04130.2 139
The following settings are possible for the display:
• brightness level of the backlit display (not in night mode)
• dimmed brightness level for energy saving (not in night mode)
• time interval after which the display switches from normal to dimmed in
the absence of operating actions or alarms (not in night mode)
• normal, night or inverted video display on the screen
The corpuls
3
never switches to AutoDIM (energy saving) mode in
defibrillation-, pacer- or nightvision mode.
If the audio recording option is enabled, a 15 second recording of the
surrounding sounds can be saved additionally by pressing the Event key (5 s
before and 10 s after pressing the key). The audio recording appears as manual
event in the protocol. Furthermore, if the screenshot option is enabled, a picture
file of an event can be saved by pressing the Event key. The audio recording
and the screenshots can be played and viewed with the software corView2 (see
chapter 8.6 Analysis of the Data with corView2, page 193).
The UTC offset (time zone) and the automatic daylight saving time (DST) have
to be configured before setting the time. The UTC and time are printed on the
protocol. For the UTC offset and the time, the unchanged minutes also have to
be confirmed with the jog dial. Only then the changed hour is saved with the
softkey [OK].
3. To confirm the settings and close the configuration dialogue, press the
softkey [OK].
To retain the previous settings and close the configuration dialogue, press
the softkey [Cancel].
Via the configuration page “Master” the master data can be edited (see chapter
8.4 Master Data, p. 188)
Audio recording
and screenshots