Configuration
ENG - Version 2.1 – P/N 04130.2 157
The actually available value range indicated in Table 7-11 depends on the
selected upper and lower limit values, as the upper and lower limits may not
overlap.
Function Lower limit Upper limit Increment
HR 1/min --; 25 to 100 --; 70 to 250 5
SpO
2
% --; 65 to 98 --; 90 to 99 1
PP 1/min --; 25 to 100 --; 70 to 235 5
SpCO % --; 0 to 99 --; 1 to 99 1
SpHb g/dl --; 5.0 to 12.0 --; 10 to 22 0,1
SpHb mmol/l --; 3.1 to 7.4 --; 6.2 to 13.7 0,1
SpMet % --; 0.1 to 99.5 --; 1 to 99.5 0.1 (0-2)
0.5 (2-100)
NIBP mmHg SYS --; 50 to 150 --; 100 to 250 5
NIBP mmHg DIA --; 10 to 80 --; 50 to 120 5
CO
2
mmHg --; 10 to 50 --; 15 to 60 1
CO
2
kPa --; 1.3 to 6.7 --; 2.0 to 8.0 0.1
RR 1/min --; 5 to 40 --; 15 to 80 1
T1 °C
--,- ; 30 to 40
*
)
--,- ; 35 to 42
*
)
0.1
T2 °C --,- ; 30 to 40 --,- ; 35 to 42 0.1
P1 to P4 mmHg SYS --; -50 to 200 --; 0 to 300
1 (- 50 to 30)
5 (30 to 300)
P1 to P4 mmHg DIA --; -50 to 200 --; 0 to 300
7-11 Values for alarm limits
*
)
For temperature measurements on the body surface, the limit values must be adjusted
accordingly.
4. To confirm the settings and close the configuration dialogue, press the
softkey [OK].
To retain the previous settings and close the configuration dialogue, press
the softkey [Cancel].