Procedure in case of malfunctions
ENG - Version 2.1 P/N 04130.2 245
Alarm message Priority Explanation/Measure
Pacer failure High priority The pacing therapy must be checked immediately.
The corpuls
3
is possibly not functioning correctly
and must not be used.
Contact authorised sales and service partners.
Pacer high impedance
High priority The electrical resistance of the patient (impedance)
is to high for the selected settings. An intensity for
stimulation has been selected that cannot be
reached with the impedance present.
Make sure that the corPatch electrodes are tightly
and completely attached to the patient's skin.
In case of excessive hair on the patient, shave the
required area on the skin. If necessary, use new
corPatch electrodes.
To perform a pacer therapy adaequate for the
patient, a higher stimulation intensity must be
selected.
Perform medical measures as needed.
Pacer short circuit
High priority
The corPatch electrodes must be placed at a
sufficient distance to allow pacing.
Make sure that the corPatch electrodes do not
touch each other.
Make sure that there is no conductive connection
between the corPatch electrodes (e.g. wetness).
Pacing not possible High priority
The corpuls
3
is possibly not functioning correctly
and must not be used.
Contact authorised sales and service partners.
Paddle interface error High priority The connectors of the therapy master cable and the
therapy electrodes have been connected while
turned the wrong way by 180 degrees and have to
be checked for damage.
If no damage is visible, connect the electrode plug
again to the therapy master cable, oriented correctly.
If the alarm message persists or damage is visible,
the corpuls
3
is not functioning correctly and must
not be used.
Contact authorised sales and service partners.
Pairing failed
Low priority The connection authorisation (Pairing) has failed.
For further information see chapter 3.2.1 Pairing
(Connection Authorisation) , page 10).
Contact authorised sales and service partners.
Phonebook update
failed
Low priority The memory location of the phonebook cannot be
written on.
If the memory location is the SIM card, check the
SIM card.
If the memory location is the corpuls
3
, contact your
authorised sales and service partners.