Procedure in case of malfunctions
ENG - Version 2.1 P/N 04130.2 249
Alarm message Priority Explanation/Measure
Temp. measurement
failed
Error T1/T2 (P-box)
Medium priority The measuring option mentioned is faulty.
Contact authorised sales and service partners.
Therapy electrodes
loose
High priority
Check all connections of the corPatch electrodes.
Connect corPatch electrodes to the patient.
Check contact of corPatch electrodes with skin.
If necessary, re-connect all connections of the
corPatch electrodes.
If the alarm message persists, connect immediately
the reserve corPatch electrodes.
Time/date invalid
Low priority The set time or the set date are invalid.
Set the correct time/date.
VT/VF possible
High priority There is possibly an arrhythmia in form of a
ventricular tachycardia (VT) or a ventricular
fibrillation (VF).
Analyse the ECG or perform an ECG interpretation
in the AED mode.
Wrong GSM PIN
Low priority The PIN belonging to the SIM card is unknown to the
corpuls
3
. The SIM card in the corpuls
3
was
possibly replaced.
Data transmission must not be repeated, to avoid
that the SIM card is locked.
Contact the persons responsible for the device.
Wrong therapy
electrodes
High priority In AED mode, there are non-suited therapy
electrodes connected to the therapy master cable.
The corpuls
3
is in pacer mode and the shock
paddles are connected.
Connect corPatch electrodes.
Table 10-1 Alarm messages, alphabetical