Procedure in case of malfunctions
User Manual corpuls
3
248 ENG - Version 2.1 P/N 04130.2
Alarm message Priority Explanation/Measure
SpCO < [NUMBER] %
SpCO > [NUMBER] %
Medium priority The measured carbonmonoxide level exceeds/falls
below the upper/lower alarm limit.
Check the patient's vital signs.
SpHb < [NUMBER] g/dl
SpHb > [NUMBER] g/dl
SpHb < [NUMBER]
mmol/l
SpHb > [NUMBER]
mmol/l
Medium priority The measured haemoglobin level exceeds/falls
below the upper/lower alarm limit.
Check the patient's vital signs.
SpMet < [NUMBER] %
SpMet > [NUMBER] %
Medium priority The measured methaemoglobin level exceeds/falls
below the upper/lower alarm limit.
Check the patient's vital signs.
SpO2 < [NUMBER] %
SpO2 > [NUMBER] %
Medium priority The measured oxygen saturation exceeds/falls
below the upper/lower alarm limit.
Check the patient's vital signs.
ß-SW-NOT FOR PAT.
USE (Defib)
ß-SW-NOT FOR PAT.
USE (Monitor)
ß-SW - NOT FOR PAT.
USE (P-Box)
High priority The software of the module is a beta version. Using
this module on a patient is prohibited.
This beta software has been released only for
testing purposes, as the functions are implemented,
but have not been tested completely.
Contact authorised sales and service partners.
SW conflict [MODULE]
High priority The connection authorisation (Pairing) has failed.
Due to different software versions the modules
cannot be connected.
For further information see chapter 3.2.1 Pairing
(Connection Authorisation) , page 10).
Contact authorised sales and service partners.
SYND CLK error
High priority The automatic synchronisation in manual
defibrillation mode does not work. No shock can be
released.
Contact authorised sales and service partners.
SYND TEST error
High priority A hardware fault exists in the defibrillator.
The corpuls
3
is not functioning correctly and must
not be used.
Contact authorised sales and service partners.
T(X) < [NUMBER] °C
T(X) > [NUMBER] °C
Medium priority „X“ stands for a temperature channel from 1 to 2.
The measured temperature exceeds/falls below the
upper/lower alarm limit.
Check the patient's vital signs.
T1/T2 sensor loose
T1/T2 sensor loose (P-
Box)
Medium priority The sensor indicated has come loose from the
measuring site at the body or from the intermediate
cable.
Check the sensor and reconnect if necessary.