Procedure in case of malfunctions
User Manual corpuls
3
270 ENG - Version 2.1 – P/N 04130.2
message line and
information in the
protocol
[NUMBER] [NUMBER]
[NUMBER] [NUMBER]
User prompt to re-enter the new access code.
Confirmation prompt asking if the access code of the user level
should be reset to factory settings.
--
To begin pacer therapy, select intensity.
Select intensity/frequency
To begin pacer therapy, select intensity and frequency.
Message in the message line in pacer mode after pressing the
softkey [Mode].
Select the operating mode of the pacer.
Status message during telemetry.
Connection to server could not be established.
Repeat attempt to establish connection.
([NUMBER] J), [NUMBER]
Ohm ([TEXT])
Time stamp and status indication of a performed defibrillation in the
protocol.
The selected energy, the effectively released energy (in brackets)
and the impedance in Ohm are indicated.
Indication of the defibrillation mode used.
Shock aborted The shock could not be delivered. Repeat shock delivery if
necessary.
The event is documented in the protocol.
The result of the ECG analysis is that a shock is not recommended.
Perform medical measures as needed. If necessary, continue with
cardio-pulmonary resuscitation.
Defibrillation energy has been released.
Check vital signs and continue cardio-pulmonary resuscitation if
necessary.
electrodes
The therapy electrodes have an electric connection (short circuit).
Make sure that the therapy electrodes do not touch.
Software conflict [MODULE]
- Separate modules
The connection authorisation (Pairing) has failed.
Due to different software versions, the modules cannot be
connected and have to be separated from each other.
For further information see chapter 3.2.1 Pairing (Connection
Authorisation) , page 10).
Contact authorised sales and service partners.
Starting time of the recorded ECG in the protocol.
Confirmation prompt asking if the stimulation should be interrupted
and the corpuls
3
should switch to manual or AED defibrillation
mode.