Procedure in case of malfunctions
User Manual corpuls
3
270 ENG - Version 2.1 P/N 04130.2
Notification in the
message line and
information in the
protocol
Explanation/Measure
Re-enter new code:
[NUMBER] [NUMBER]
[NUMBER] [NUMBER]
User prompt to re-enter the new access code.
Reset code?
Confirmation prompt asking if the access code of the user level
should be reset to factory settings.
Select Energy
--
Select frequency
--
Select intensity
To begin pacer therapy, select intensity.
Select intensity/frequency
To begin pacer therapy, select intensity and frequency.
Select mode
Message in the message line in pacer mode after pressing the
softkey [Mode].
Select the operating mode of the pacer.
Server disconnected
Status message during telemetry.
Connection to server could not be established.
Repeat attempt to establish connection.
Shock [NUMBER] J
([NUMBER] J), [NUMBER]
Ohm ([TEXT])
Time stamp and status indication of a performed defibrillation in the
protocol.
The selected energy, the effectively released energy (in brackets)
and the impedance in Ohm are indicated.
Indication of the defibrillation mode used.
Shock aborted The shock could not be delivered. Repeat shock delivery if
necessary.
The event is documented in the protocol.
Shock not recommended
The result of the ECG analysis is that a shock is not recommended.
Perform medical measures as needed. If necessary, continue with
cardio-pulmonary resuscitation.
Shock performed
Defibrillation energy has been released.
Check vital signs and continue cardio-pulmonary resuscitation if
necessary.
Short circuit - Check
electrodes
The therapy electrodes have an electric connection (short circuit).
Make sure that the therapy electrodes do not touch.
Software conflict [MODULE]
- Separate modules
The connection authorisation (Pairing) has failed.
Due to different software versions, the modules cannot be
connected and have to be separated from each other.
For further information see chapter 3.2.1 Pairing (Connection
Authorisation) , page 10).
Contact authorised sales and service partners.
Start ECG
Starting time of the recorded ECG in the protocol.
Switch off pacer?
Confirmation prompt asking if the stimulation should be interrupted
and the corpuls
3
should switch to manual or AED defibrillation
mode.