Operation – Therapy
User Manual corpuls
3
86 ENG - Version 2.1 – P/N 04130.2
The pacing intensity can be adjusted in 5 mA increments within the range of
0 to 150 mA.
WARNING
Regularly check effectiveness of the pacer by checking the central pulse.
WARNING
In patients with an implanted pacer, it is possible that shockable ECG rhythms
or arrhythmias will only be detected to a limited extent.
Fig. 5-14 Pacer, selecting the intensity
If necessary, press the softkey [Pause] to interrupt pacing. Answer the
confirmation prompt Pause pacing? with the softkey [Yes].
If pacing has been paused, press the softkey [Continue pacing] to continue
pacing. Answer the confirmation prompt Continue pacing? with the softkey
[Yes].
To end the current pacing, press the softkey [Off]. Answer the confirmation
prompt Switch off pacer? with the softkey [Yes] to end pacing and to reset
the pacer (DEMAND, 0 mA, 70/min).
WARNING
The procedure described below is a recommendation from the manufacturer.
Qualified users will determine the progress of treatment on their own account.