198
Error Indication
<Internal Memory Error>
Code Error indication Cause Corrective action/more information
ME01 Internal memory is full.
Delete les or reformat.
The instrument’s internal
memory is full.
Perform this procedure.
1. If recording is in progress, stop
recording.
2. Copy the data on the internal memory to
the SD memory card (p. 125).
3. Delete all the les on the internal
memory (p. 125) or format the internal
memory (p. 126).
ME02 Cannot be used because
internal memory is
damaged.
Proceed with
reformatting?
YES: ENTER key
No: ESC key
The instrument’s internal
memory is corrupt.
Format the internal memory (p. 126).
<Operation Error>
Code Error indication Cause Corrective action/more information
OP01 This folder cannot be
deleted.
You attempted to delete
the [PQ3100] base folder.
The [PQ3100] base folder cannot be deleted.
If you wish to delete it, you must do so on a
computer.
<Error>
Code Error indication Cause Corrective action/more information
ER01 Invalid setting value. You attempted to congure
the setting with a value that
is outside the valid setting
range.
Congure the setting with a value that falls
within the valid setting range.
See “5 Setting Change (SET UP Screen)”
(p. 63)
ER02 Maximum number of
recordable events
exceeded.
The number of events
exceeds the upper limit,
9999. No further events
can be recorded.
Stop the recording and change the event
threshold value in order that 10000 events
do not occur.
See “5 Event Settings” (p. 72)