60
Checking Wiring
If the wiring judgment result is red (FAIL) or yellow (CHECK)
1
Select an item to be checked.
Move the cursor
A dialog box with helpful information for xing
the wiring will be displayed.
2
Review its contents.
(Press the [ESC] key to close the dialog box)
3
Press the [WIRING] key to display the
WIRING, wiring settings screen.
4
Verify that the actual wiring connections
are the same as the ones shown on the
screen.
Correct the wiring if the connections are
incorrect.
5
Once again, press [WIRING] key and
conrm the wiring judgment result on
WIRING, wiring check conrmation
screen.
Recording can be started even if the wiring
judgment result remains red (FAIL) or yellow
(CHECK).