Start and Stop of the Recording
Start recording at an appropriate time punctuation
Recording start: interval, recording stop: manual or set time
Example: Recording interval is set to 10 min
Present time
10:41:22
Recording start
10:50:00
WAITING RECORDING
If the recording interval is 30 sec or below, the recording will start at “10:42:00.”
Repeat recording
Recording start: repeat, recording stop: repeat, recording period: variable period
Example 1: If the recording period is 00:00 to 24:00 and recording interval is 10 min.
Present time
10:41:22
Recording start
10:50:00
Recording (electric
energy) reset
24:00:00
Recording resume
00:00:00
Recording (electric
energy) reset
24:00:00
Recording resume
00:00:00
WAITING
RECORDING RECORDING
RECORDING
(Repeat until
the stop date)
Example 2: If the
recording period
is
08:00 to 18:00
Recording start
08:00
WAITING
Start date
Recording start
08:00
WAITING
Recording stop
18:00
RECORDING
RECORDING
Recording stop
18:00
(Repeat until the stop date)
Present time
7:32
If the set time has already passed when the [START/STOP] key is pressed, the recording will start at an
appropriate time punctuation (Interval).

Table of Contents

Other manuals for Hioki PQ3100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hioki PQ3100 and is the answer not in the manual?

Hioki PQ3100 Specifications

General IconGeneral
BrandHioki
ModelPQ3100
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals