EasyManua.ls Logo

Hioki PQ3100 - Recording Start Timing Options; Repeat Recording Setup Procedures

Hioki PQ3100
242 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
92
Start and Stop of the Recording
Start recording at an appropriate time punctuation
Recording start: interval, recording stop: manual or set time
Example: Recording interval is set to 10 min
Present time
10:41:22
Recording start
10:50:00
WAITING RECORDING
If the recording interval is 30 sec or below, the recording will start at “10:42:00.”
Repeat recording
Recording start: repeat, recording stop: repeat, recording period: variable period
Example 1: If the recording period is 00:00 to 24:00 and recording interval is 10 min.
Present time
10:41:22
Recording start
10:50:00
Recording (electric
energy) reset
24:00:00
Recording resume
00:00:00
Recording (electric
energy) reset
24:00:00
Recording resume
00:00:00
WAITING
RECORDING RECORDING
RECORDING
(Repeat until
the stop date)
Example 2: If the
recording period
is
08:00 to 18:00
Recording start
08:00
WAITING
Start date
Recording start
08:00
WAITING
Recording stop
18:00
RECORDING
RECORDING
Recording stop
18:00
(Repeat until the stop date)
Present time
7:32
If the set time has already passed when the [START/STOP] key is pressed, the recording will start at an
appropriate time punctuation (Interval).

Table of Contents

Other manuals for Hioki PQ3100

Related product manuals