16
Part Names and Functions
Key operation area
MENU keys (switching screens)
Pressing a key causes the selected key to light up and the screen to switch over to the selected screen.
Displays the Measurement screen.
The Measurement screen displays measured values and waveforms.
p. 51
Displays the Input Settings screen.
The Input Settings screen is used to congure settings related to input, wiring
mode, measurement, and calculations.
p. 41
Displays the System Settings screen.
The System Settings screen is used to congure settings related to time control,
interfaces, and other functionality.
p. 115
Displays the File Operation screen.
The File Operation screen is used to handle les.
p. 119
Channel indicators
The display settings of the RANGE key and the setting indicator a󰀨ect the input channels that are lit up.
The channels included in a single wiring conguration based on the wiring settings light up simultaneously.
Channel selection key
Selects the channel to display on the Measurement screen.
The channel indicators light up linking to the channel selection key.
RANGE keys
The + and keys of U can switch the voltage range, while the + and keys of I
can switch the current range.
These a󰀨ect the ranges of the channel with it channel indicator LED lit up.
When the [A-D] channel indicator is lit up, the U keys functions for Ch. A analog
input, while the I keys functions for Ch. C analog input.
When the [E-H] indicator is lit up, the U keys functions for Ch. E analog input,
while the I keys functions for Ch. G analog input.
When the AUTO key is lit up, auto-ranging operation is canceled at the time
when the range is switched.
AUTO keys
The AUTO key in the U area enables the auto-ranging function for voltage,
while the AUTO key in the I area enables the auto-ranging function for current.
The keys will light up. The key light turns o󰀨 if pressed again, and the range will
be xed to the present setting at that time.
The keys function for the channels with their channel indicators are lit up.
Performs zero adjustment for the input channels. p. 46
Saves measured data onto the USB ash drive when the key is pressed. p. 119
Saves the screen image onto the USB ash drive when the key is pressed. p. 132
REMOTE/LOCAL key (key lock)
The key lights up when the instrument enters the remote state during the GP-IB
communications. The key light turns o󰀨 if it is pressed again and the instrument
returns to the local state.
If the key is held down for at least 3 s, the key lock is activated and the key lock
icon is displayed on the screen. If the key is held down again for at least 3 s,
the setting is released and the key light turns o󰀨.
p. 173
HIOKI PW8001A961-00

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hioki PW8001-01 and is the answer not in the manual?

Hioki PW8001-01 Specifications

General IconGeneral
BrandHioki
ModelPW8001-01
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals