EasyManua.ls Logo

Honeywell DR4500A - Table 5-18 Procedure for Running a Setpoint Ramp

Honeywell DR4500A
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation - Single Setpoint Ramp
134 DR4500A Classic Series Circular Chart Recorder With or Without Control Product Manual Release M
April 2017
Running the setpoint ramp
Running a setpoint ramp includes starting, holding, viewing the ramp time, ending the ramp and disabling
it.
Procedure
76HTable 5-18 lists the procedure for running the setpoint ramp.
Table 5-18 Procedure for running a setpoint ramp
Step
Operation
Press
Action/Result
1
Put the controller into
automatic mode
MAN
AUTO
Until “A” indicator is ON and you will see:
Upper Display: H and the PV value
Lower Display: SP and the present setpoint value
2
Set Start SP
or
Until the start setpoint value you desire is indicated in
the lower display:
Upper Display: H and the PV value
Lower Display: SP and the start setpoint value
3
Start the ramp
RUN
HOLD
You will see:
Upper Display: R and the PV value
Lower Display: SP and a changing setpoint value
NOTE: The value in the lower display will be increasing
or decreasing toward the final setpoint value.
The PV value in the upper display and the pen trace will
also change.
4
Hold the ramp at the
current value
RUN
HOLD
This holds the ramp at the current setpoint value.
Press again to continue run.
A “KEY ERROR” prompt appears if RUN/HOLD key is
pressed while “SP RAMP” is disabled.
5
View the remaining ramp
time
LOWR
DISP
Until you see:
Upper Display: R or H and the PV value
Lower Display: RAMP XXXM (time remaining)
6
End the ramp
When the final setpoint is reached, the “R” changes to
“H” in the upper display and the recorder operates at
the new setpoint.
ATTENTION
Any time the local setpoint is different
from the final setpoint value and the RUN/HOLD key is
pressed, the ramp will start again.

Table of Contents

Related product manuals