EasyManuals Logo

Honeywell DR4500A User Manual

Honeywell DR4500A
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #199 background imageLoading...
Page #199 background image
Input Calibration - Calibration Set Up and Wiring for PV Inputs
Release M DR4500A Classic Series Circular Chart Recorder With or Without Control Product Manual 185
April 2017
Thermocouple inputs using a compensated calibrator method
Refer to Figure 7-4 and wire the recorder according to the procedure given in Table 7-7.
Table 7-7 Set up wiring procedure for Thermocouple inputs using a compensated
calibrator method
Step
Action
1
Connect the thermocouple extension wire to the Precision Compensated Calibrator (see
Figure 7-4).
2
Connect the other end of the thermocouple extension wires to the clamp type terminals in the
plug for the applicable input connector (J2).
3
Reinstall the plug into the input connector (J2) on the applicable circuit board for Input 1 or
Input 2.
#2 #1
22090
Connector plug J2 f or
inputs #1 or #2
+
_
Thermocouple
extension wires
_
+
Compensated
Calibrator
MA
W1
Be sure W1/MA
jumper is in position
W1 on the input board.
T/C
RTD
Be sure W3 jumper
is in position T/C
on the input board.
Figure 7-4 Calibration set up diagram for Thermocouple inputs using a compensated
calibrator method

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell DR4500A and is the answer not in the manual?

Honeywell DR4500A Specifications

General IconGeneral
BrandHoneywell
ModelDR4500A
CategoryRecording Equipment
LanguageEnglish

Related product manuals