EasyManuals Logo

Honeywell DR4500A User Manual

Honeywell DR4500A
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #204 background imageLoading...
Page #204 background image
Input Calibration - Calibration Set Up and Wiring for PV Inputs
190 DR4500A Classic Series Circular Chart Recorder With or Without Control Product Manual Release M
April 2017
420 mA inputs
Refer to Figure 7-9 and wire the recorder according to the procedure given in Table 7-12.
Table 7-12 Set up wiring procedure for 420 mA inputs
Step
Action
1
Connect the copper wire to the calibration source (see Figure 7-9).
2
Connect the other end of the copper wire to the clamp type terminals in the plug for the
applicable input connector (J2).
3
Reinstall the plug into the input connector (J2) on the applicable circuit board for input 1 or
input 2.
#2 #1
22305
+
_
_
+
Copper wires
Calibration source
Plug J2
MA
W1
Be sure W1/MA jumper is in
position MA; this connects
250 resistor (internal on
PCB) across inputs terminals
4-20 mA
T/C
RTD
Be sure W3 jumper
is in position T/C
on the input board.
Figure 7-9 Calibration set up diagram for 420 mA inputs

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell DR4500A and is the answer not in the manual?

Honeywell DR4500A Specifications

General IconGeneral
BrandHoneywell
ModelDR4500A
CategoryRecording Equipment
LanguageEnglish

Related product manuals