EasyManuals Logo

Honeywell DR4500A User Manual

Honeywell DR4500A
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #235 background imageLoading...
Page #235 background image
Troubleshooting / Service
Pen Alignment
Release M DR4500A Classic Series Circular Chart Recorder With or Without Control Product Manual 221
April 2017
Table 9-20 Set the chart time line procedure
Step
Press
Action/Result
1
Open the door, apply power, and wait for the recorder to run its power up
tests.
2
SET
UP
Until you see:
Upper Display: ALIGN
Lower Display: PEN 1
3
FUNC
You will see:
Upper Display: ALIGN
Lower Display: PEN 10
The pen moves to an internal stop then backs away and stops.
â–² or â–¼
To move the pen to the zero position on the chart (inner position of chart).
ATTENTION: You can position the pen over 100 steps. Those steps are
relative units only. Step 0.0 is the innermost position on the chart, and step
1200 is the highest circumference position.
4
FUNC
You will see:
Upper Display: ALIGN
Lower Display: PEN 100
The value adjusted in Step 3 is entered into memory.
â–² or â–¼
To raise or lower the value in the upper display (up to 1200) move the pen
to the highest chart range value (outer circumference printed on the chart).
5
Repeat steps 2 through 4 for pen 2 or press LOWR DISP key to return to
normal operation and enter the span value into memory.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell DR4500A and is the answer not in the manual?

Honeywell DR4500A Specifications

General IconGeneral
BrandHoneywell
ModelDR4500A
CategoryRecording Equipment
LanguageEnglish

Related product manuals